中国激光, 1998, 25 (10): 891, 网络出版: 2006-10-18
关于小角前向散射激光测粒仪准则数X的讨论
A Discussion on the Criterion Number X of a Small angle Forward Scattering Laser Particle Sizer
激光测粒仪 小角前向散射 FD理论 Mie散射理论 准则数X laser particle sizer small angle forward scattering FD theory Mie scattering theory criterion number X
摘要
在激光测粒仪的设计和数据处理过程中,都采用了无因次准则数x=1.357。之后,为了降低激光测粒仪的测量下限,经典的Mie散射理论在激光测粒仪的设计中得到了应用。讨论了在利用Mie光散射理论时,无因次准则数x的取值及其影响因素。
Abstract
The criterion number X which equals 1.357 is adopted in the design and data processing of the laser FD based particle sizes. The classical Mie light scattering theory is used in the design of the laser particle sizers to extend their lower size limit. According to the theory, the criterion number X and the factors influencing X are discussed in detail.
沈建琪, 王乃宁. 关于小角前向散射激光测粒仪准则数X的讨论[J]. 中国激光, 1998, 25(10): 891. 沈建琪, 王乃宁. A Discussion on the Criterion Number X of a Small angle Forward Scattering Laser Particle Sizer[J]. Chinese Journal of Lasers, 1998, 25(10): 891.