光学学报, 2006, 26 (8): 1226, 网络出版: 2007-03-15   

医用硬性内窥镜畸变的评定基础和方法

Evaluation Base and Method of Medical Rigid Endoscope Distortion
作者单位
国家食品药品监督管理局杭州医疗器械检验中心, 杭州 310009
摘要
建立医用硬性内窥镜畸变的评定基础和方法,用视轴对称的球面Z视场为模型来模拟体腔凹面特征,以建立内窥镜物像共轭的分析基础;运用光学原理并结合ISO9039:1994“光学和光学仪器光学系统质量评价——畸变的测定”的定义原则,并根据理论畸变的临床要求确定畸变定义;采用主光线追迹原理推导不同球面Z形状的畸变换算通式。创立了医用硬性内窥镜以单位相对畸变表述的定义和表达式,并导出变换视场面的畸变换算通用公式和测试方法。本文建立的畸变评定基础和方法适用于医用硬性内窥镜的标准应用。
Abstract
To set up the evaluation base and method of medical rigid endoscope distortion, by using optical axis symmetrical spherical Z view field as the model to simulate concave characteristics of body cavity, the analysis base of endoscope object-image conjugation is established. By using optical principle together with the definition principle in ISO 9039:1994 “Optics and optical instruments——Quality evaluation of optical systems——Determination of distortion”, with definition of distortion determined by clinical requirement of theoretical distortion, the distortion conversion formula of different sphere Z shapes is deduced by main ray trace principle. The medical rigid endoscope distortion definition and expression which are expressed as unit relative distortion, and distortion conversion formula and testing method of conversion of view field surface, are all established. The distortion evaluation base and method established are applicable for the medical rigid endoscope standard.

贾晓航, 颜青来, 文燕. 医用硬性内窥镜畸变的评定基础和方法[J]. 光学学报, 2006, 26(8): 1226. 贾晓航, 颜青来, 文燕. Evaluation Base and Method of Medical Rigid Endoscope Distortion[J]. Acta Optica Sinica, 2006, 26(8): 1226.

本文已被 5 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!