中国激光, 2008, 35 (9): 1429, 网络出版: 2008-09-09   

光纤型光学相干层析技术系统的眼科成像

Ophthalmic Imaging by Optical Coherence Tomography
作者单位
1 中国科学院光电技术研究所, 四川 成都 610209
2 中国科学院研究生院, 北京 100039
3 浙江大学现代光学仪器国家重点实验室, 浙江 杭州 310027
摘要
建立了一套单模光纤型光学相干层析(OCT)成像系统,开展了动物眼睛的成像实验,实现了高信噪比、高分辨率、大成像深度的层析图像的获取。系统纵向分辨率达到9 μm,横向分辨率为10 μm,最大层析成像深度可达3.4 mm。实验图像和商业蔡氏三代光学相干层析技术成像的对比结果表明,研制的OCT系统已经能展示视网膜的基本分层结构,而且能分辨脉络膜的分层结构和脉络膜血管,后者是蔡氏三代OCT系统所无法实现的。
Abstract
Single-mode fiber-designed optical coherence tomography (OCT) system is developed and applied in ophthalmic imaging. Images with high signal-to-noise ratio, high resolution and large image depth, are achieved by the developed system. The calibrated axial resolution and transversal resolution are 9 μm and 10 μm, respectively, and the maximum depth range is 3.4 mm. Comparison between experimental results and Zeiss OCT Ⅲ are presented. The results show that the proposes OCT system can image not only the layers of retina , but also the layers and vas of choroid which cannot be obtained by Zeiss OCT Ⅲ.

史国华, 丁志华, 戴云, 饶学军, 张雨东. 光纤型光学相干层析技术系统的眼科成像[J]. 中国激光, 2008, 35(9): 1429. Shi Guohua, Ding Zhihua, Dai Yun, Rao Xuejun, Zhang Yudong. Ophthalmic Imaging by Optical Coherence Tomography[J]. Chinese Journal of Lasers, 2008, 35(9): 1429.

本文已被 14 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!