光学学报, 2009, 29 (10): 2813, 网络出版: 2009-10-19   

单色仪转角重复精度的相位板衍射准直方法测量

Measurement of Monochromator′s Angle Repeatability by Phase Plate Diffraction Collimation Method
作者单位
中国科学院 上海应用物理研究所,上海 201800
摘要
为准确测量上海光源软X射线谱学显微光束线采用的变包含角平面光栅单色仪的转角重复精度,提出了一种新的基于相位板衍射准直技术的测量方法。该方法将半导体激光单模光纤和相位板衍射准直技术结合起来,利用面阵CCD采集图像,通过测量光斑的位移变化确定平面镜和光栅的角度变化。实验表明,该方法可以测量掠入射情况下单色仪联动时的转角重复精度,测量精度可达±0.1″,此测量精度优于同等实验条件下的商用ELCOMAT 3000自准直仪的测量精度。
Abstract
In order to measure the mechanical angle repeatability of the variable-included-angle plane-grating-monochromator used in soft X-ray spectro-microscopy beamline at Shanghai Synchrotron Radiation Facility,a new method was proposed based on the phase plate diffraction collimation. This system was designed combining laser diode fiber with phase plate diffraction collimation technique. A 2-dimension CCD was adopted to collect the images,which was used to measure the displacement of the light spot,thus the angle repeatability of the plane mirror and the grating could be determined. The results show that this method could be used to measure the monochromator′s angle repeatability in a higher precision estimated to be ±0.1″,which is better than that measured by the commercial ELCOMAT 3000 auto collimator used in the same condition.

巩志华, 陈敏, 王勇, 严睿, 邰仁忠. 单色仪转角重复精度的相位板衍射准直方法测量[J]. 光学学报, 2009, 29(10): 2813. Gong Zhihua, Chen Min, Wang Yong, Yan Rui, Tai Renzhong. Measurement of Monochromator′s Angle Repeatability by Phase Plate Diffraction Collimation Method[J]. Acta Optica Sinica, 2009, 29(10): 2813.

本文已被 3 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!