光电工程, 2008, 35 (6): 42, 网络出版: 2010-03-01   

紫外辐射计的波长定标及不确定度分析

Wavelength Calibration for Ultraviolet Spectroradiometer and Analysis of its Uncertainty
作者单位
1 中国科学院长春光学精密机械与物理研究所应用光学国家重点实验室,长春 130031
2 中国科学院研究生院,北京 100039
摘要
限于常规汞灯谱线法波长定标的局限性,构建了紫外辐射计波长定标装置,研究了紫外辐射计波长定标的物理过程和测量链,并对紫外辐射计中臭氧十二个吸收波长进行了光谱定标,通过对定标影响量的分析和计算,得到定标影响量的测量不确定度和波长定标合成标准不确定度,其中波长定标合成标准不确定度为0.026nm,同时通过光学CAD分析和实验验证紫外辐射计的光谱带宽可以达到1.0nm。应用自行构建的紫外波长定标装置较好地完成了紫外辐射计臭氧吸收谱线的定标工作,满足了臭氧反演所需的波长精度要求和光谱带宽要求。
Abstract
Referring to limitation of the conventional mercury lamp spectral calibration in ultraviolet wave band, a spectral calibration device was built for ultraviolet spectroradiometer. The physical process and measurement chain of spectral calibration were studied and twelve ozone absorptive spectral lines were calibrated. The measurement uncertainty and combined standard uncertainty of influenced quantities of wavelength calibration were analyzed and the combined standard uncertainty was 0.026nm. Spectral bandwidths can reach 1.0nm according to the analysis of optical CAD and experiment results. The spectral calibration system does well in ultraviolet spectral calibration for spectroradiometer. The calibrated results satisfy the spectral precision for ozone retrieval.

王加朋, 王淑荣, 徐领娣, 宋克非, 林雪松, 黄煜, 汪龙祺, 王锐, 张振铎, 李福田. 紫外辐射计的波长定标及不确定度分析[J]. 光电工程, 2008, 35(6): 42. WANG Jia-peng, WANG Shu-rong, XU Ling-di, SONG Ke-fei, LIN Xue-song, HUANG Yu, WANG Long-qi, WANG Rui, ZHANG Zhen-duo, LI Fu-tian. Wavelength Calibration for Ultraviolet Spectroradiometer and Analysis of its Uncertainty[J]. Opto-Electronic Engineering, 2008, 35(6): 42.

本文已被 5 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!