光子学报, 2009, 38 (6): 1491, 网络出版: 2010-05-10   

人眼视网膜成像自适应光学系统的初步试验和改进

Primary Experiment and Improvement Design of Adaptive Optics System for Human Retinal Imaging
作者单位
1 中国科学院长春光学精密机械与物理研究所,应用光学国家重点实验室,长春 130033
2 中国科学院研究生院,北京 100049
摘要
搭建了一套基于液晶空间光调制器的人眼视网膜成像自适应光学系统,进行了活体人眼视网膜的初步实验.经过系统闭环校正,PV值和RMS值分别从2.293λ降低到0.176 553λ,从0.55129λ降低到0.105 11λ,接近衍射极限的水平.获得了较为清晰的人眼视网膜细胞图像,验证了液晶空间光调制器在人眼视网膜高分辨率自适应成像中应用的可行性,并针对试验中的遇到的激光散斑以及照明控制等问题,对原系统提出了一些改进设计.
Abstract
An adaptive optics retinal imaging system based on liquid crystal spatial modulator was set up and a experiment on lively human eye was made.After adaptive optics correction,the PV (peak to valley) and the RMS(root mean square) decreased from 2.293λ to 0.176 553λ,from 0.551 29λ to 0.105 11λ respectively,and the high-resolution retinal image was obtain.Thus it is proved that liquid crystal spatial modulator can be applied into the high-resolution retinal imaging.Some improving design was also presented according to the practical problem encountering in experiment such as laser speckle and illumination control.

程少园, 胡立发, 曹召良, 穆全全, 姜宝光, 李鹏飞, 宣丽. 人眼视网膜成像自适应光学系统的初步试验和改进[J]. 光子学报, 2009, 38(6): 1491. CHENG Shao-yuan, HU Li-fa, CAO Zhao-liang, MU Quan-quan, JIANG Bao-guang, LI Peng-fei, XUAN Li. Primary Experiment and Improvement Design of Adaptive Optics System for Human Retinal Imaging[J]. ACTA PHOTONICA SINICA, 2009, 38(6): 1491.