摘要
为了实现皮秒拍瓦激光的精密测试,研制了专用的皮秒自相关仪。选用无色散元件,提高了测量结果的可靠性。采用自标定方法标定时间分辨率,用超限法标定时间测量范围,用极值法标定脉冲响应特性。实验结果表明,皮秒自相关仪的时间分辨率为0.1 ps/pixel,时间测量范围为26 ps,脉冲响应特性的值为300 fs。该皮秒自相关仪能够满足皮秒啁啾脉冲的时间宽度精密测试要求。
Abstract
In order to realize precise diagnostics for pico-second pulse of peta-watt laser, an auto-correlator especially for single shot pico-second pulse is developed. Optical elements with no dispersion are used in this pico-second auto-correlator′s design, which could improve its precise. Three methods are adopted to testify its characters. Self-calibration is used for time resolution. Over-range is used for time range. Limitation is used for impulse response function. Experiments results show that its time resolution is 0.1 ps/pixel, its time range is 26 ps, and its impulse response function value is 300 fs. So this auto-correlator can fulfill the requirement of precised measurement on pulse width for chirped pulse of pico-second peta-watt laser.

欧阳小平, 杨琳, 彭永华, 唐顺兴, 刘崇, 马金贵, 钱列加, 冷雨欣, 朱宝强, 朱健强. 皮秒自相关仪的性能测试研究[J]. 中国激光, 2012, 39(4): 0408003. Ouyang Xiaoping, Yang Lin, Peng Yonghua, Tang Shunxing, Liu Chong, Ma Jingui, Qian Liejia, Leng Yuxin, Zhu Baoqiang, Zhu Jianqiang. Research on Precise Diagnostics for Pico-Sesond Single Shot Pulse[J]. Chinese Journal of Lasers, 2012, 39(4): 0408003.

本文已被 7 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!