强激光与粒子束, 2012, 24 (8): 1871, 网络出版: 2012-08-09   

化学气相沉积金刚石X射线探测器相对标定

Relative calibration of chemical vapor deposited diamond X-ray detector
作者单位
中国工程物理研究院 激光聚变研究中心, 四川 绵阳 621900
摘要
利用神光Ⅱ激光装置,开展了化学气相沉积金刚石X射线探测器的相对标定技术研究。实验得到了金刚石X射线探测器与已绝对标定的平响应X射线二极管对X射线辐射的测量结果,计算得到金刚石X射线探测器平均灵敏度为1.196×10-5 C/J,不同发次得到的灵敏度与平均值之间的偏差不大于13%。
Abstract
The relative calibration of a chemical vapor deposited (CVD) diamond X-ray detector has been completed on the shenguang-Ⅱ laser facility. Signals of the CVD diamond detector and an absolute-calibrated flat response X-ray diode (XRD) for X-ray radiation were obtained. The average sensitivity of the CVD diamond X-ray detector is about 1.196×10-5 C/J. The difference between average sensitivity and the sensitivities in different shots is no more than 13%.

侯立飞, 李志超, 袁永腾, 况龙钰, 杨国洪, 刘慎业. 化学气相沉积金刚石X射线探测器相对标定[J]. 强激光与粒子束, 2012, 24(8): 1871. Hou Lifei, Li Zhichao, Yuan Yongteng, Kuang Longyu, Yang Guohong, Liu Shenye. Relative calibration of chemical vapor deposited diamond X-ray detector[J]. High Power Laser and Particle Beams, 2012, 24(8): 1871.

本文已被 2 篇论文引用