光谱学与光谱分析, 2013, 33 (12): 3269, 网络出版: 2014-01-09  

基于可调谐多模二极管激光吸收光谱的二氧化碳浓度测量

Study on CO2 Measurement Using Tunable Multi-Mode Diode Laser Absorption Spectroscopy
作者单位
1 上海理工大学光电信息与计算机工程学院, 上海 200093
2 滨州学院物理与电子科学系, 理论物理研究所, 山东 滨州 256603
3 教育部光学仪器与系统工程研究中心, 上海市现代光学系统重点实验室, 上海 200093
摘要
可调谐二极管激光吸收光谱技术(TDLAS)具有高分辨率、 高灵敏度和快速测量等特点, 是气体探测中的主流技术。 使用多模二极管激光器作为光源, 将多模二极管激光吸收光谱技术与关联光谱技术相结合有助于提高测试可靠性和稳定性, 同时还可有效解决单模二极管激光器长时间工作时中心波长随外界温度或机械特性的变化而发生偏移的问题。 以1 570 nm多模二极管激光器为光源, 利用多模二极管激光关联光谱和波长调制的气体测量技术(TMDL-COSPEC-WMS), 通过计算待测气体和参考池气体之间二次谐波信号峰值高度之间的关系, 实现了对二氧化碳浓度的测量。 实验中二氧化碳浓度测量范围在0.6%~30%之间, 计算结果表明, 二氧化碳浓度与真实浓度值之间具有良好的线性关系, 其线性度为0.998 7, 线性拟合的斜率为1.061±0.016 8。 对二氧化碳与氮气混合气体的连续测量结果表明, 系统的探测极限达到335 ppm·m, 对同一样品在20 min内的20次连续测量的标准偏差为0.036 7%, 表明了系统良好的稳定性, 所有测量结果都显示了系统用于二氧化碳气体监测的有效性。
Abstract
Tunable diode laser absorption spectroscopy (TDLAS) technology is a kind of fast time response, large-range, continuous on-line monitoring gas detection technique. It is the mainstream technology of gas detection. In this paper the multimode laser diode was used as light source. Multi-mode laser combined with correlation spectroscopy can improve the test reliability and stability. It can also conquer the problem of the central wavelength change of the single mode diode laser due to thermal or mechanical fluctuations in durable working process. A FP laser was used as the light source in this research. A multi-mode diode laser system based on correlation spectroscopy and wavelength modulation spectroscopy (TMDL-COSPEC-WMS) was used to measure carbon dioxide in ambient air around 1 570 nm. The carbon dioxide concentrations were derived from the relationship between the normalized WMS-2f signal peak heights of the measurement and reference signals which selected based on high signal to noise ratio and correlation coefficient. All measurements were performed with controlled carbon dioxide and nitrogen mixtures in which carbon dioxide concentrations range from 0.6% to 30%. The calculation results showed that there was a high linear relationship between the measured and actual carbon dioxide concentration, the linearity was 0.998 7 and the fitted slope was 1.061±0.016 8 respectively over the tested range. A detection limit of 335 ppm·m was achieved. The standard deviation of 0.036 7% was achieved using 20 successive measurements with each measurement time taking ~10 s during 20 minutes, which demonstrated good stability of the system. Good agreements between the measurements of the system and actual values confirm the accuracy and potential utility of the system for carbon dioxide detection.

高光珍, 陈抱雪, 胡波, 龙秀慧, 李爱萍, 李荣. 基于可调谐多模二极管激光吸收光谱的二氧化碳浓度测量[J]. 光谱学与光谱分析, 2013, 33(12): 3269. GAO Guang-zhen, CHEN Bao-xue, HU Bo, LONG Xiu-hui, LI Ai-ping, LI Rong. Study on CO2 Measurement Using Tunable Multi-Mode Diode Laser Absorption Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2013, 33(12): 3269.

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!