激光与光电子学进展, 2014, 51 (3): 031204, 网络出版: 2014-03-03   

基于泰伯-莫尔干涉技术的焦距测量系统校准技术研究 下载: 517次

Calibration of Foci-Meter Based on Talbot-Moiré Interferometer
作者单位
1 西南技术物理研究所, 四川 成都 610041
2 南京理工大学电光学院, 江苏 南京 210000
3 四川大学电子信息学院, 四川 成都 610041
摘要
研制了一套用于测量超长焦距的泰伯-莫尔干涉仪,并设计了一套校准设备和方法对泰伯-莫尔干涉仪的测量不确定度进行评定,核准焦距范围达100 m。校准设备包括了一套焦距发生器和激光光源,利用经过计量的标准透镜和实验对焦距发生器进行标定、测量和量值溯源。通过实验结果分析和计算,可知该校准技术具有良好的测量不确定度,能够对基于泰伯-莫尔方法的焦距测量系统进行精确校准。
Abstract
It developed a method to calibrate the foci- meter based on Talbot- Moiré interferometer, the calibration range of focal length up to 100 meters. Moreover, it introduced a foci-meter based on Talbot-Moiré interferometer. It includes a lens group and a laser source to generate laser beams with steady focal length and wavefronts which used to calibrate the foci- meter. The focal length of the lens group is measured by experiments taking advantage of a traceable lens.The following conclusions can be derived from experimental results: the uncertainty of this calibration method is acceptable.

赵琦, 蒋泽伟, 孟庆安, 李轶国, 樊红英, 胡绍云, 李建欣, 何勇, 杨洁. 基于泰伯-莫尔干涉技术的焦距测量系统校准技术研究[J]. 激光与光电子学进展, 2014, 51(3): 031204. Zhao Qi, Jiang Zewei, Meng Qing′an, Li Yiguo, Fan Hongying, Hu Shaoyun, Li Jianxin, He Yong, Yang Jie. Calibration of Foci-Meter Based on Talbot-Moiré Interferometer[J]. Laser & Optoelectronics Progress, 2014, 51(3): 031204.

本文已被 1 篇论文引用