光学与光电技术, 2014, 12 (2): 79, 网络出版: 2014-05-04
甚长波宽波段(6.4~15 μm)红外探测器响应光谱特性的测量
Measurement of Response Spectral Characteristics for Long Wavelength Broad-Band (6.4~15 μm) Infrared Detectors
傅里叶变换光谱仪 甚长波宽波段红外探测器 探测器饱和 响应光谱 响应时间 Fourier transform spectrometer long wavelength broad-band (6.4~15 μm) infrared d detector saturation response spectrum response time
摘要
使用傅里叶变换光谱仪(FTIR)测试甚长波宽波段(6.4~15 μm)红外探测器响应光谱的过程中,发现短波方向响应光谱异常。通过分步测试分析发现:探测器和放大器工作在非线性工作区导致某些情况下仪器信号发生饱和,引起了短波方向响应光谱畸变的现象。对FTIR测量甚长波宽波段(6.4~15 μm)红外探测器响应光谱的畸变现象进行了分析,认为探测器的响应时间是影响其响应光谱的重要因素,并通过试验确定了测试系统对不同探测器所设置的测试参数,消除了响应光谱畸变的现象,并提高了测试准确度。
Abstract
In measuring response spectrum of long wavelength broad-band (6.4~15 μm) infrared detectors by using Fourier Transform Spectrometer(FTIR), the response spectral distortion at short-wave is found. Through the test step by step. It is discovered that the spectrum signal will saturate if the detectors and amplifiers work in non-linearity workspace, resulting in the phenomenon of spectral distortion at short-wave. In this article, the signal saturation is investigated, response time is an important factor for response spectrum of different detectors. FTIR measurement for different detectors test parameters are studied. The phenomenon of response spectral distortion is eliminated, and the measurement accuracy is improved by using the appropriate parameters.
贺香荣, 张亚妮, 汪洋. 甚长波宽波段(6.4~15 μm)红外探测器响应光谱特性的测量[J]. 光学与光电技术, 2014, 12(2): 79. HE Xiang-rong, ZHANG Ya-ni, WANG Yang. Measurement of Response Spectral Characteristics for Long Wavelength Broad-Band (6.4~15 μm) Infrared Detectors[J]. OPTICS & OPTOELECTRONIC TECHNOLOGY, 2014, 12(2): 79.