光学仪器, 2014, 36 (2): 112, 网络出版: 2014-05-22   

光栅单色仪中光谱重建的影响因素分析

Research on influencing factors of spectra reconstruction in grating monochromator
作者单位
1 北京信息科技大学 光电测试技术北京市重点实验室,北京100192
2 北京信息科技大学 光电信息与仪器北京市工程研究中心,北京100192
摘要
针对一种以Ⅳ型全息凹面光栅为核心元件的单色仪,为了保证仪器的光谱重建精度以提高波长准确性,通过频谱计算和理论推导对光谱重建过程中的抽样间距、光谱半高宽、探测器的光敏单元宽度和灵敏度常数等影响因素进行了分析,讨论了它们的取值关系对系统能否满足抽样定理的影响,得到了抽样间距与光敏单元宽度的合理取值区域,并应用在该型单色仪的设计实例中,为同类结构的单色仪提供了准确重建光谱的思路与方法。
Abstract
Focusing on a kind of monochromator which uses type Ⅳ holographic concave grating as core element, in order to ensure the precision of spectra reconstruction and accuracy of wavelength, through spectrum calculating and theoretical derivation, this paper analyses the influencing factors of spectra reconstruction in the instrument, such as integral interval, full width half maximum of spectra line, photosensitive width and sensitivity constant of detector and so on. Through discussing about the influence of these factors on the system, whether it could fulfill the sampling theorem, this paper gets the reasonable value area of integral interval and photosensitive width and application instances in the design of monochromator and provides a kind of thought and method about accurate spectra reconstruction for similar monochromator.

许梦泽, 郭阳宽, 祝连庆, 娄小平, 孟晓辰. 光栅单色仪中光谱重建的影响因素分析[J]. 光学仪器, 2014, 36(2): 112. XU Mengze, GUO Yangkuan, ZHU Lianqing, LOU Xiaoping, MENG Xiaochen. Research on influencing factors of spectra reconstruction in grating monochromator[J]. Optical Instruments, 2014, 36(2): 112.

本文已被 2 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!