中国激光, 2015, 42 (3): 0315002, 网络出版: 2015-02-13   

激光诱导击穿光谱技术检测铀材料中微量杂质元素

Detection of Trace Impurities in Uranium Using Laser Induced Breakdown Spectroscopy
作者单位
表面物理与化学重点实验室, 四川 绵阳 621907
摘要
研究了利用激光诱导击穿光谱技术对铀材料中杂质元素进行快速定量分析的方法。利用KrF 准分子激光器和AvaSpec-2048 光纤光谱仪组建了激光诱导击穿光谱系统,检测并分析了200~460 nm 波长范围内铀及杂质元素的光谱数据。分别研究了脉冲激光次数、光谱仪检测延迟时间等因素对铀光谱特性的影响。在激光能量为28 mJ、延迟时间为1.5 μs 时对铀材料中的Fe、Cu、Al 三种杂质元素进行定量分析,特征光谱分别选择438.3、427.5、396.2 nm ,定量测量相对误差小于12%。实验结果表明激光诱导击穿光谱技术在检测铀材料中微量杂质元素时具有较低的相对误差和检出限,可以用于铀材料中杂质元素的快速定量分析。
Abstract
Feasibility of rapid quantitative analysis for trace impurities in uranium materials is studied using laser induced breakdown spectroscopy (LIBS). LIBS system is set up with a KrF excimer laser and an AvaSpec-2048 fiber optic spectrometer. Spectral data both of uranium and impurity elements are detected and analyzed in a range of 200~460 nm wavelength. The influence of pulse number and delay time of spectrometer detection on the characteristics of uranium and impurity elements is investigated respectively. Three kinds of impurity elements Fe, Cu and Al in uranium are quantitatively analyzed with characteristic spectra at 438.3, 427.5, 396.2 nm respectively. The relative error of quantitative measurement is less than 12% when the laser energy is 28 mJ and time delay is 1.5 μs . The experimental results show that laser induced breakdown spectroscopy which has relatively lower relative error and limit of detection can be used for rapid quantitative analysis of trace impurity elements in uranium materials.

徐钦英, 张永彬, 王怀胜, 贾建平. 激光诱导击穿光谱技术检测铀材料中微量杂质元素[J]. 中国激光, 2015, 42(3): 0315002. Xu Qinying, Zhang Yongbin, Wang Huaisheng, Jia Jianping. Detection of Trace Impurities in Uranium Using Laser Induced Breakdown Spectroscopy[J]. Chinese Journal of Lasers, 2015, 42(3): 0315002.

本文已被 12 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote