强激光与粒子束, 2015, 27 (3): 032022, 网络出版: 2015-03-23   

单光子计数闪烁探测器灵敏度标定方法

Sensitivity calibration of NaI(Tl) scintillator based on single photon counting
作者单位
1 中国工程物理研究院 激光聚变研究中心, 四川 绵阳 621900
2 中国原子能科学研究院, 北京 100049
摘要
根据NaI(Tl)闪烁探测器可用于单光子计数的特性, 探索了一种测量放射源入射下闪烁探测器脉冲电荷谱来实现其绝对灵敏度标定的方法。实验对标定系统的电荷数字转换记录器进行了绝对标定, 得到了用于ICF实验中重要诊断设备滤波-荧光谱仪的NaI(Tl)闪烁探测器的绝对灵敏度。灵敏度标定数据不确定度小于10%, 较传统的放射源标定方法(15%)显著提高, 同时标定能量范围扩大到100 keV以上。最后, 实验结果与Geant4模拟程序计算的结果进行了比对, 符合得较好。
Abstract
This paper mainly introduced some researches about calibration of scintillators sensitivity using standard radio -active sources 241Am, 133Ba and 152Eu in Beijing Institute of Atomic Physics. As a scintillator could be used for single photon counting, are developed a new method to calibrate a scintillator’s sensitivity by measuring detectors’ charge spectrum with radioactive sources’ incidence. The experiment also calibrated QDC(quantity-to-digital convertor), and got the results of sensitivity for Filter-Fluorescer spectrometer in ICF experiments. The data’s uncertainty (10%) was lower than that with traditional methods (15%), and calibration energy range increased up to more than 100 keV. Finally, we compared experimental data with Geant4 simulating results, and they were in good accordance.

何小安, 李朝光, 易荣清, 苏明, 张奇玮, 阮锡超. 单光子计数闪烁探测器灵敏度标定方法[J]. 强激光与粒子束, 2015, 27(3): 032022. He Xiaoan, Li Chaoguang, Yi Rongqing, Su Ming, Zhang Qiwei, Ruan Xichao. Sensitivity calibration of NaI(Tl) scintillator based on single photon counting[J]. High Power Laser and Particle Beams, 2015, 27(3): 032022.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!