光学学报, 2015, 35 (6): 0630004, 网络出版: 2015-05-20   

大口径光谱系统狭缝弯曲的分析与矫正

Analysis and Correction of Slit Curvature in Spectrograph with High NA
作者单位
华中科技大学武汉光电国家实验室, 湖北 武汉 430074
摘要
随着大口径光谱仪的发展,传统光谱仪中并不明显的狭缝弯曲越来越成为制约光谱仪分辨率的一大因素。根据由离轴光线光栅方程获得的偏移量公式对狭缝弯曲进行矫正。为消除公式与实际图像不契合的问题,详细分析了各装配误差对公式的影响,总结出相关矫正参数,并利用光谱线丰富的光源拟合出相关预矫正参数来消除装配误差造成的影响,再结合偏移量公式对待处理光谱图像进行矫正。利用氖光谱灯拟合预矫正参数,对氪光谱灯与汞蒸汽荧光灯进行了光谱测试,经矫正后光谱分辨率有很大改善,接近于光谱仪点源照射时的分辨率。
Abstract
Slit curvature becomes an important factor which introduces the decline of spectral resolution in the developing of spectrograph with high NA. Slit curvature is corrected according to the correction formula obtained from the grating equation of off- axis ray. To cope with the problem that the real image of spectral lines usually deviate from the formula at ideal condition, calibration with a spectral lamp which has lines distributed across the full span of the spectrograph is used to fit out the assembly error parameters. Neon lamp is used to fit out these parameters, slit curvature in the spectral image of krypton lamp and fluorescent lamp of mercury vapor is corrected with these parameters. Resolution has be significantly improved, which is nearly close to the spectral resolution limit of the spectrograph when irradiated by a spot light source.

李哲, 余安澜, 左都罗, 王新兵. 大口径光谱系统狭缝弯曲的分析与矫正[J]. 光学学报, 2015, 35(6): 0630004. Li Zhe, Yu Anlan, Zuo Duluo, Wang Xinbing. Analysis and Correction of Slit Curvature in Spectrograph with High NA[J]. Acta Optica Sinica, 2015, 35(6): 0630004.

本文已被 3 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!