激光与光电子学进展, 2015, 52 (9): 093001, 网络出版: 2015-08-28   

激光诱导击穿光谱单变量及多元线性回归方法研究 下载: 580次

Research on Univariate and Multiple Linear Regression Calibration Methods by Laser Induced Breakdown Spectroscopy
作者单位
中国科学院安徽光学精密机械研究所中国科学院环境光学与技术重点实验室, 安徽 合肥 230031
摘要
为了提高激光诱导击穿光谱技术的检测能力和准确性,分别采用单变量分析和多变量分析方法[多元线性回归(MLR)]对样品中Cr 进行定量分析。分别利用CrI: 425.435 nm 和CrI: 427.48 nm 两条特征谱线进行单变量分析,并在5 种不同激光能量下获得了Cr 的检测限,结果表明谱线CrI: 425.435 nm 的分析结果要优于CrI: 427.48 nm,获得Cr 最佳的检测限为5.8 mg/g。利用多变量分析方法,研究了浓度预测值与浓度实际值之间的线性相关性,与单变量分析方法相比,线性相关性由0.98 提高到了0.99 以上;采用留一交叉验证方法,比较了两种方法的预测相对误差(REP),单变量分析方法的REP 分别为6.73%和7.59%,而MLR 分析方法的REP 则为4.66%,结果表明激光诱导击穿光谱技术结合MLR 能够提高样品浓度预测的准确性。
Abstract
In order to improve the detection capability and accuracy of laser-induced breakdown spectroscopy (LIBS), the univariate and multivariate method [multiple linear regression (MLR)] are adopted to make quantitative analysis on Cr contained in samples respectively. Univariate analysis has been made by use of two characteristic spectral lines CrI: 425.435 nm and CrI: 427.48 nm, the limits of detection are obtained under the condition of five laser energies. The results indicate that the analytical results of CrI: 425.435 nm are better than CrI: 427.48 nm, the best limit of detection of Cr is achieved 5.8 mg/g. The linear correlation between the prediction and actual values of concentration of samples is discussed using MLR, the correlation coefficient is increasing to above 0.99 from 0.98 compared with univariate method. Using the approach of leave-one-cross-validation, the relative errors of prediction (REP) of univariate analysis of CrI: 425.435 nm and CrI: 427.48 nm are 6.73% and 7.59% respectively, while the REP of MLR is reduced to 4.66% . The results indicate that LIBS combined with MLR can improve the accuracy of prediction on concentration of samples.

余洋, 赵南京, 王寅, 方丽, 孟德硕, 胡丽, 谷艳红, 王园园, 马明俊, 刘建国, 刘文清. 激光诱导击穿光谱单变量及多元线性回归方法研究[J]. 激光与光电子学进展, 2015, 52(9): 093001. Yu Yang, Zhao Nanjing, Wang Yin, Fang Li, Meng Deshuo, Hu Li, Gu Yanhong, Wang Yuanyuan, Ma Mingjun, Liu Jianguo, Liu Wenqing. Research on Univariate and Multiple Linear Regression Calibration Methods by Laser Induced Breakdown Spectroscopy[J]. Laser & Optoelectronics Progress, 2015, 52(9): 093001.

本文已被 4 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!