Photonic Sensors, 2015, 5 (3): 278–283, Published Online: Oct. 22, 2015  

Large Dynamic Range SPR Measurements Using a ZnSe Prism

Author Affiliations
interdisciplinary Photonics Laboratories, School of Chemistry, The University of Sydney, NSW 2006, Australia
Abstract
A large dynamic index measurement range (n =1 to n = 1.7) using surface plasmon resonance (SPR) shifts was demonstrated with a ZnSe prism at 632.8 nm, limited by the available high index liquid hosts. In contrast to borosilicate based SPR measurements, where angular limitations restrict solvent use to water and require considerable care dealing with Fresnel reflections, the ZnSe approach allows SPR spectroscopies to be applied to a varied range of solvents. An uncertainty in angular resolution between 1.5° and 6°, depending on the solvent and SPR angle, was estimated. The refractive index change for a given glucose concentration in water was measured to be n = (0.114 ± 0.007) /%[C6H12O6]. Given the transmission properties of ZnSe, the processes can be readily extended into the mid infrared.

John CANNING, Jing QIAN, Kevin COOK. Large Dynamic Range SPR Measurements Using a ZnSe Prism[J]. Photonic Sensors, 2015, 5(3): 278–283.