气压对激光诱导击穿光谱检测性能的影响研究

Investigation of the Effects of Atmospheric Pressure on the Performance of Laser-Induced Breakdown Spectroscopy
刘莉 *
作者单位
宜春学院物理科学与工程技术学院,江西 宜春 336000
摘要
为研究气压对激光诱导击穿光谱检测性能的影响,实验研究了0.01~100.00 kPa大气压强下钢铁样品中Fe I 438.35 nm和Fe I 404.58 nm谱线的强度、寿命、信噪比及相对标准偏差随采集时间的演化。结果表明,大气压强从100.00 kPa逐渐下降至0.01 kPa时,大气与等离子体的相互作用减小,导致等离子体光谱强度逐渐减小至原来的1/20,谱线寿命从常规的几十微秒逐渐缩短至只有2.5 μs;谱线的信噪比随气压降低逐渐减小,相对标准偏差在气压降低至0.1 kPa以下时迅速增大。这说明气压降低会导致光谱质量变差,为保证分析灵敏度,低气压下应选择较短的采集时延;当气压低于0.10 kPa时,分析谱线的相对标准偏差迅速大于10%。
Abstract
To investigate the effects of atmospheric pressure on the performance of laser-induced breakdown spectroscopy, thetemporal evolution of the intensities, lifetimes, signal-to-noise ratiosand relative standard deviation of Fe I 438.35 nm and Fe I 404.58 nm spectral lines were experimentally analyzed. The results show that, when the pressures are decreased from 100.00 kPa to 0.01 kPa, the intensities are reduced to 1/20 of the original values; the lifetimes reduce gradually from tens of microseconds to only 2.5 μs, while the signal-to-noise ratios gradually reduce and relative standard deviations increases.This indicates that a reduction in pressure can cause deterioration of the spectrum quality, to ensure the analysis sensitivity a shorter acquisition time delay should be selected for a low pressure; when the pressure is lower than 0.10 kPa, the relative standard deviations of the lines begin to exceed 10% quickly.

刘莉. 气压对激光诱导击穿光谱检测性能的影响研究[J]. 应用激光, 2015, 35(5): 588. Liu Li. Investigation of the Effects of Atmospheric Pressure on the Performance of Laser-Induced Breakdown Spectroscopy[J]. APPLIED LASER, 2015, 35(5): 588.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!