光子学报, 2016, 45 (6): 0604001, 网络出版: 2016-07-26   

光子计数器的线性测量和修正

Measurement and Correction of Linearity of Photon Counters
作者单位
1 中国科学技术大学 环境科学与光电技术学院,合肥 230026
2 中国科学院安徽光学精密机械研究所 通用光学定标与表征技术重点实验室,合肥 230031
摘要
依据自发参量下转换过程中泵浦光功率和相关光子的线性关系,搭建了光子计数器线性测量实验装置,分析了光子计数器非线性误差产生的机理,探讨其线性应用条件和修正方法.通过调节泵浦光功率改变入射至光子计数器的相关光子速率,实现了光子计数器线性的测量,并对实验测量值进行了修正.结果表明,通过线性测量和修正可以提高单光子计数器测量准确度,降低光子计数器的非线性误差,其入射光子速率最大线性工作范围可由105s-1扩大到107s-1.
Abstract
Linear measurement device was set up based on the linear relationship between pump power and photons in spontaneous parametric down conversion process. The nonlinearity errors mechanism, linear condition and correction model of photon counting detector were analyzed. For the sake of realizing the linear measurement of the photon counting, the pump power was adjusted to change incident photon rate, and the experimental measurements were corrected. The results show that the measurement precision can be improved by measurement and correction. The maximum linear range rate of the incident photon could be expanded from 105/s to 107/s result from lower non-linear errors.

胡友勃, 李健军, 夏茂鹏, 高冬阳, 郑小兵. 光子计数器的线性测量和修正[J]. 光子学报, 2016, 45(6): 0604001. HU You-bo, LI Jian-jun, XIA Mao-peng, GAO Dong-yang, ZHENG Xiao-bing. Measurement and Correction of Linearity of Photon Counters[J]. ACTA PHOTONICA SINICA, 2016, 45(6): 0604001.

本文已被 6 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!