二维相关光谱分析在无创血糖检测特异性研究中的应用

Application of Two-Dimensional Near-Infrared Correlation Spectroscopy in the Specificity Analysis of Noninvasive Blood Glucose Sensing
作者单位
1 天津大学精密测试技术及仪器国家重点实验室, 天津 300072
2 天津大学精密仪器与光电子工程学院, 天津 300072
摘要
主要利用二维相关光谱分析方法考察了血液中主要成分、 测量过程中的系统漂移等因素对近红外光谱中的葡萄糖分子特异性的影响。 首先测量了葡萄糖水溶液的近红外透射光谱, 以葡萄糖浓度为外扰, 计算二维相关同步谱及异步谱, 通过对比分析确定了葡萄糖在合频和倍频区域的特征吸收峰位置。 当葡萄糖水溶液中加入少量白蛋白后, 其二维相关谱中葡萄糖的一级倍频吸收峰(1 590 nm)和二级倍频吸收峰(1 195 nm)出现了不同源的现象, 表明白蛋白可能破坏了葡萄糖的特异性; 进一步开展了人体口服葡萄糖耐量实验, 测量了手掌部位的漫反射光谱, 其二维相关谱中葡萄糖的同源性吸收也遭到破坏。 另外, 由于葡萄糖浓度变化引起的光谱变异信息较小, 很容易被系统漂移掩盖, 一般需要进行背景修正。 在葡萄糖水溶液的透射实验和人体漫反射实验中, 分别引入纯水样本和5%标准漫反射板作为参考, 以时间为扰动的二维相关光谱分析的结果表明, 经过参考样本的背景校正后, 系统漂移在二维同步切线谱中引起的谱峰偏移变小, 葡萄糖的特征吸收被增强。 因此, 在近红外光谱无创血糖检测中, 应尽量避开与葡萄糖特征吸收不同源的波段, 并有必要采取一定的参考测量方法以提高检测的特异性。
Abstract
In this paper, two-dimensional (2D) correlation spectroscopy analysis was applied to investigate the influence of the main component in blood and the systematic drift during the measurement on the specificity of glucose in the near-infrared (NIR) spectroscopy. First, the NIR transmittance of glucose aqueous solutions was measured and the 2D correlation NIR spectra were calculated under the perturbation of glucose concentration. Based on the comparative analysis for synchronous and asynchronous 2D correlation spectra, the characteristic absorption peaks of glucose in the combination band and the overtone band were determined. Then a small amount of albumin was added into glucose aqueous solutions, and the transmittance was recorded to perform 2D correlation spectroscopy analysis under the perturbation of glucose concentration. However, the absorption of glucose in the first overtone band (1590nm) and second overtone band (1195nm) was no longer homologous in the 2D correlation spectra, which means that the albumin may reduce the specificity of glucose. Further, the oral glucose tolerance test of healthy volunteer was conducted and the NIR diffuse reflectance of left palm was collected in vivo. The 2D correlation analysis results showed that, the homology of glucose in the diffuse reflectance was also destroyed. Moreover, as the spectral variation from the glucose concentration change is too low to be covered by that induced by systematic drift easily, some background correction methods were usually required. For the transmittance experiment of glucose aqueous solutions and the diffuse reflectance experiment of human body, the pure water sample and 5% diffuse reflectance standard were used as the reference, respectively. Then 2D correlation spectroscopy was developed under the perturbation of measurement time. Results showed that, smaller band shift was observed in the slice spectra of 2D correlation synchronous spectra after the corresponding background correction, and the specificity of glucose was improved both in the in vitro and in vivo experiments. So for the non-invasive glucose sensing by NIR spectroscopy, the wavelengths should be chosen carefully to avoid the absorption band of some interfering components which may destroy the homology of glucose and make spectral interpretation more complicated. And the selection of reference samples for relative measurement is also important to improve the specificity of glucose.

胡永翔, 刘蓉, 张雯, 徐可欣. 二维相关光谱分析在无创血糖检测特异性研究中的应用[J]. 光谱学与光谱分析, 2017, 37(2): 491. HU Yong-xiang, LIU Rong, ZHANG Wen, XU Ke-xin. Application of Two-Dimensional Near-Infrared Correlation Spectroscopy in the Specificity Analysis of Noninvasive Blood Glucose Sensing[J]. Spectroscopy and Spectral Analysis, 2017, 37(2): 491.

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!