Chinese Optics Letters, 2017, 15 (10): 100101, Published Online: Jul. 19, 2018   

Development and application of automated vicarious calibration system Download: 921次

Author Affiliations
1 Key Laboratory of Optical Calibration and Characterization, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China
2 University of Sciences and Technology of China, Hefei 230026, China
3 Troop 61741 of PLA, Beijing 100094, China
4 Shanghai Maritime University, Shanghai 201306, China
Abstract
In order to improve the frequency and precision of radiometric calibration, the automated vicarious calibration system (AVCS) is developed and deployed at the Dunhuang test site to perform vicarious calibration without the in situ manned measurements. The surface and atmospheric parameters are automatically collected by AVCS. An absolute radiometric calibration approach based on AVCS is proposed. Six successful calibrations of the Aqua Moderate Resolution Imaging Spectroradiometer (MODIS) are conducted. The results are in good agreement with the on-board calibration system with all the relative differences less than 4%. It enables us to monitor the change of a sensor over long time scales.

Wei Wei, Shuai Song, Yangang Sun, Ganggang Qiu, Xin Li, Xiaobing Zheng. Development and application of automated vicarious calibration system[J]. Chinese Optics Letters, 2017, 15(10): 100101.