光电子快报(英文版), 2017, 13 (6): 436, Published Online: Sep. 13, 2018   

The nonlinearity measurement of charge-coupled device array spectrometer using colorful LED

Author Affiliations
National Institute of Metrology, Beijing 100029, China
Abstract
A nonlinearity measurement of the charge-coupled device (CCD) array spectrometer using flux addition and comparison method is described. The light with various colors from the colorful light emitting diode (LED) light source is applied to measure the nonlinearity of the spectrometer at different wavelengths, respectively. An high-end CCD array spectrometer is tested. For colorful LED light sources, the nonlinearity factors of the CCD array spectrometer (absolute value) are as follows: k<0.8% for white light, k <1.1% for red light, k <2.2% for green light and k<4.7% for blue light. By using those quasi-monochromatic light sources, it is shown that the nonlinearity depends on the wavelength. It is important to be wariness about the spectral nonlinearity and related uncertainty evaluation when the narrow-band light source is tested.

ZHAO Wei-qiang, LIU Hui, LIU Jian. The nonlinearity measurement of charge-coupled device array spectrometer using colorful LED[J]. 光电子快报(英文版), 2017, 13(6): 436.

本文已被 1 篇论文引用
被引统计数据来源于中国光学期刊网