光谱学与光谱分析, 2018, 38 (3): 697, 网络出版: 2018-04-09   

基于微型多次反射腔的TDLAS二氧化碳测量系统

Study on TDLAS System with a Miniature Multi-Pass Cavity for CO2 Measurements
作者单位
中国海洋大学光学光电子实验室, 山东 青岛 266100
摘要
海气界面CO2测量对于海洋科学研究具有重要意义, 在目前的海洋CO2测量仪器中, 基于可调谐二极管激光吸收光谱技术(TDLAS)的设备因灵敏度高、 环境适应性强等特点受到关注。 TDLAS系统的体积和灵敏度通常受限于多次反射腔的大小和光程。 针对海洋CO2脱气量小且灵敏度高的测量需求, 自主设计了一套微型多次反射腔, 用于TDLAS系统的CO2测量。 该微型多次反射腔采用两片口径为25.4 mm、 焦距为50 mm的球面反射镜, 以38 mm的腔长实现了253次反射, 获得了约10 m的光程, 封装后的样品池体积仅有90 mL。 基于该微型多次反射腔搭建了一套直接吸收TDLAS的CO2气体浓度测量系统, 通过标准气体对该系统进行了测试, 检测限约为26×10-6, 不同浓度气体线性相关度R2为99.986%。 同时还将该系统与LGR公司生产的便携式温室气体分析仪(UGGA)进行了对比测量, 结果表明二者在白天CO2浓度波动较大和夜晚CO2浓度变化较平稳两种情况下均表现出较好的一致性, R2大于97%。 实验结果证明了系统性能, 下一步将优化试验装置并进行现场应用。
Abstract
The measurement of CO2 at air-sea interface is significant to marine scientific research. There are many commercial instruments can be used for CO2 measurements in marine environment, among which the instruments based on tunable diode laser absorption spectroscopy (TDLAS) play an important role due to the advantages of high sensitivity, good environmental adaptability, etc. As a core component of a TDLAS system, the volume and optical path of a multi-pass cavity limit the size and sensitivity of the system. For the application of underwater dissolved CO2 measurements, due to the samll quantity of degassing gas and high sensitivity, both the small size and long optical path are required for a multi-pass cavity. In this work, a miniature multi-pass cavity was designed for the potential underwater dissolved CO2 measurements. The miniature multi-pass cavity composed of two identical spherical mirrors (D=25.4 mm, f=50 mm)separated at a distance of 38 mm, realized 253 reflection times, providing an effective optical path length of 10 m and an inner volume of 90 mL. Based on the miniature multi-pass cavity, a direct absorption TDLAS system was developed for CO2 measurement. The system was evaluated with a series of different concentrations of CO2 standard gases. The obtained limit of detection (LOD) was about 26×10-6 (volume ratio), and the response was good linear with R2=99.986% over the whole range. A commercial ultraportable greenhouse gas analyzer (UGGA) from Los Gatos Research (LGR) was also used for comparison measurements, and the results showed a consistent trend with R2 of more than 97% under conditions of both high fluctuation at daytime and low changes at nighttime in laboratory. The experiments testified the performance of the developed TDLAS system for CO2 measurements, which means that the improved system will be used in field experiments in the near future.

李萌, 郭金家, 叶旺全, 李楠, 张志浩, 郑荣儿. 基于微型多次反射腔的TDLAS二氧化碳测量系统[J]. 光谱学与光谱分析, 2018, 38(3): 697. LI Meng, GUO Jin-jia, YE Wang-quan, LI Nan, ZHANG Zhi-hao, ZHENG Rong-er. Study on TDLAS System with a Miniature Multi-Pass Cavity for CO2 Measurements[J]. Spectroscopy and Spectral Analysis, 2018, 38(3): 697.

本文已被 4 篇论文引用
被引统计数据来源于中国光学期刊网
引用该论文: TXT   |   EndNote

相关论文

加载中...

关于本站 Cookie 的使用提示

中国光学期刊网使用基于 cookie 的技术来更好地为您提供各项服务,点击此处了解我们的隐私策略。 如您需继续使用本网站,请您授权我们使用本地 cookie 来保存部分信息。
全站搜索
您最值得信赖的光电行业旗舰网络服务平台!