Author Affiliations
Abstract
1 College of Information Engineering, Southwest University of Science and Technology, Mianyang 621010
2 College of Precision Instrument, Tianjin University, Tianjin 300072
An improved scattering optical model was developed under cylindrical coordinate to simulate the thermal effect of diffusing applicator in laser interstitial thermotherapy (LITT). The thermal damage was calculated by finite element method (FEM) using Pennes bio-heat transfer equation and Arrhenius injury integral formula. The numerical results showed that the scattering can considerably influence the evaluation of the lesion area, and the relationship between application powers or time and resulting tissue thermal damage was nonlinear. Although usually applying relatively low power can avoid tissue charring, rather higher power is recommended because it is indispensable to achieve necessary damage threshold and the therapy time can be shortened.
医用光学 生物光学 数值模拟 光热效应 激光热损伤 000.4430 Numerical approximation and analysis 350.5340 Photothermal effects 140.3330 Laser damage 